MEA1060-Up-Standard

Probe interface and integrated 60 channel pre- and filter amplifier for 60 electrode MEAs (microelectrode arrays), for upright microscopes

  • MEA1060-Up-Standard amplifier:
    Gain and bandwidth as specified on the order.
    (Standard amplifier specifications:
    Gain: 1200, bandwidth: 1 Hz - 3 kHz;
    possible amplifier specifications:
    Gain: 500-5000, bandwidth: 1 Hz - 10kHz)
    Direct access to each electrode for stimulation via connectors
    Integrated heating element and temperature sensor
  • C68x3M:
    68-pin MCS standard cable, 3 m
  • Accessories:
    Cable assembly, reference electrodes, test model probe, spare parts, user manual
  • Technical specifications as specified in the data sheet unless noted otherwise in the order.
 Datasheet for MEA1060-Up-Standard