MEA60-Up-BC-System-E-Standard

Complete recording system for micro-electrode arrays on upright microscopes including 1 amplifier to record from 60 MEA electrodes and 4 additional channels. With enhanced perfusion equipment and a two channel temperature controller. Featuring a blanking circuit for simultaneous stimulation and recording without stimulus artefacts.

  • Data acquisition computer:
    PC configuration 2, with preinstalled hardware, operating system, and software
  • MC_Rack Program:
    Data acquisition and analysis software, including a 2 years maintenance package (software updates and support)
  • MEA_Select Program:
    MEA1060-BC control software, including a 2 year maintenance package (software updates and support)
  • MC_Card 64:
    64-channel PCI-bus data acquisition card, simultaneous 50 kHz sampling on each channel
  • 68-Pin-Slot:
    Interface for 64 analog inputs and power distribution (68-pin MCS Standard connector)
  • 3-An-Slot:
    Interface for 3 analog inputs (3 BNC connectors)
  • 3-Di-Slot:
    Interface for input bits 0, 1, 2 of the 16-bit digital (TTL) channel (3 BNC connectors)
  • IPS10W:
    Low-noise isolated power supply with 10 W power and ± 6.3 V output voltage
  • MEA1060-Up-BC-Standard amplifier (incl. MEA1060-Up-BC-PA, FA60SBC and C68x1M):
    Probe interface and integrated 60 channel pre- and filter amplifier for 60 electrode MEAs, for inverted microscopes, with blanking circuit, controlled by MEA_Select software.
    Gain and bandwidth as specified in the order.
    (Standard amplifier specifications:
    Gain: 1100, bandwidth: 1 Hz - 3 kHz;
    possible amplifier specifications:
    Gain: 500-5000, bandwidth: 1 Hz - 10kHz)
  • C68x2M:
    68-pin MCS standard cable, 2 m
  • TC02:
    2-channel temperature controller for biological samples, set-point temperature reached fast within 30 s to 5 minutes, control temperature range from ambient temperature to +50 °C,
    precision controller using PID characteristics
  • PH01:
    Perfusion cannula with programmable fluid temperature, controlled by TC01/2
  • 5 MEAs:
    Type and electrode grid as specified in the order (complete catalog range available)
  • Accessories:
    Cable assembly, reference electrodes, test model probe, spare parts, user manuals


Technical specifications as specified in the data sheet unless noted otherwise in the order.