www.multichannelsystems.com

Innovations in Electrophysiology

MEA2100-Beta-Screen-2-System

Manufacturer: 
Multi Channel Systems MCS GmbH
Quality Made in Germany

In-vitro MEA system for diabetes research with beta cells in intact islets of Langerhans.

Compact data acquisition system for up to eight MEA2100-Beta-Screen headstages with 5 channels each. Complete with two MEA2100-beta-screen-headstage, MCS-signal collector unit, MCS-interface board 3.0 multiboot, data acquisition and analysis software, two Beta-Screen microelectrode arrays, integrated temperature controllers, and constant vacuum pumps, power supply, and accessories. USB Super Speed data transfer with a sampling rate of up to 50 kHz/channel.

  • Data acquisition and analysis software Beta-Screen
    - Including a 2 years maintenance package (software updates and support)

  • MCS interface board 3.0 multiboot (MCS-IFB 3.0 Multiboot):
    Includes various in-/outputs:
    - 4x Digital Out: Interface for output bit 0-3 of the 16-bit digital input / output channels (Lemo connector)
    - 4x Digital In for synchronization with other instruments (bit 0-3) (Lemo connector)
    - Digital IN / OUT: Interface for 16-bit digital input / output channels. Generates or accepts TTL-pulses
    - 1x Stereo Out: Possibility to export 2 channels (e.g on an oscilloscope or headphones making the signals audible)
    - 2x Analog IN: Interface for 2 additional analog inputs (channel 1 and 2) (Lemo connector)
    - Analog IN: 8-channel Analog input (channel 1-8)
    - 1x Ground
    - 1x Universal Serial Bus Super Speed for transferring digitally converted data to any data acquisition computer with a sampling rate of up to 50 kHz (USB Super Speed cable)
    - 1x USB Serial Bus Super Speed for programing the integrated signal processor
    - External power supply: 100-240 V input voltage range

  • MCS signal collector unit (MCS-SCU):
    - Output for analog data to interface with 3rd party data acquisitions
    - Four programmable outputs for optical stimulation (optogenetics)
    - Up to two signal collector units can be connected to one MCS-IFB

  • 2x MEA2100-Beta-Screen-HS
    Compact 5 channel headstage for MEA2100-Beta-Screen-System.
    - Fully digital 5 channel headstage with built-in FPGAs
    - Bandwidth adjustable via software
    - Data resolution: 24 bit
    - Sampling rate max. 50 kHz per channel
    - Integrated heating device, controlled by the Beta-Screen software
    - Includes fluidic ports for perfusion and suction port to adhere clusters to the electrodes

  • 2x Beta-Screen-MEA
    Microelectrode array for cell cluster measurements:
    - TiN electrodes
    - PI isolator
    - With internal reference electrode
    - Linear electrode arrangement
    - 5 electrodes
    - Electrode spacing 570 μm
    - Electrode diameter 50 μm
    - perforated Polyimide foil
    - Integrated heating element
    - Warranty: 6 months

  • 2x PPS2
    Peristaltic perfusion system with one inlet and one outlet pump.
    -  Dimensions H x W x D: 235 x 160 x 110 mm
    - Weight: 3.45 kg
    - Flow rate freely adjustable via touch screen, software (included) or analog inputs
    - Flow rate inlet pump: 0 ml/min - 30 ml/min
    - Flow rate drain pump: 0 ml/min - 30 ml/min
    - Low pulsation pump design
    - Additional droplet isolation chambers to further absorb pulsation
    - Long life brushless motor drives
    - Low electromagnetic emission
    - Constant flow rates even at lowest setting
    - Supply voltage: 100VAC to 240VAC
    - Manual
    - 2 sets of replacement tubes incl. droplet isolation chambers

  • 2x CVP-230V
    Constant vacuum pump with pressure control.
    - Dimensions H x D x W: 0.47 x 0.15 x 0.23 m
    - Weight: 7.4 kg
    - Resolution of pressure control: 0.1 mbar
    - Maximum negative pressure: -200 mbar
    - Accuray of the sensor: +/- 2.5 %
    - Supply voltage: 230 VAC
    - Power supply
    - Manual

  • Accessories:
    - Spare parts
    - USB memory stick with user manuals and software

Technical specifications as specified in the data sheet unless noted otherwise in the order.