www.multichannelsystems.com

Innovations in Electrophysiology

W2100-HS32-opto

Manufacturer: 
Multi Channel Systems MCS GmbH
Quality Made in Germany

Wireless headstage with 32 recording channels with interface for optical stimulation and control of applied current.
32-channel headstage for W2100-System for recording from 32 in vivo electrodes. Wireless transmission of digitized data from headstage to receiver.

  • 1 x Headstage:
      - Dimensions (W x D x H): 15.5 (+antenna) x 15.5 x 9.2 mm (+ battery)
      - Weight: 4.8 g (+ battery)
      - 32-channel data acquisition
      - Bandwidth: 1 Hz – 5 kHz (0.1 Hz on request)
      - Resolution: 16 bit
      - Input voltage range: +/- 12.4 mV
      - Input noise: < 1.9 µV RMS
      - Sampling rate: 20 kHz/channel on 32 to 16 selected channels, 25 kHz/channel 15 to 5 selected channels, 40 kHz/channel on 4 or less selected channels
      - 2 LEDs (status adjustable via software)
      - With triaxial gyroscope sensor
      - With triaxial accelerometer sensor
      - Electrode connector: Omnetics A79023
      - Control of applied stimulus current (up to 1 A per channel)
      - Interface for connection of LED for optical stimulation (LED not included; connector type Mill-Max 1mm pitch female)

Technical specification as specified in the data sheet. Subject to change without notice.

References